System type |
Automated bench top clinical chemistry analyzer |
|
STAT-analytics |
2 sample positions available |
|
Throughput |
150 tests/hour for mono-reagents; 100 tests/hour for 2 component
reagents
|
|
Ion measurement |
Enzymatic tests for Cl, K, Na |
|
Bar code identification |
Automatic bar code scan for reagents and samples |
|
Measuring principle |
Colorimetry (Rate/End Point); Turbidimetric Immunoassay
|
|
Calibration |
Linear, non-linear, multi-point |
|
Combined reagent/sample tray |
30 reagent positions plus 30 patient sample positions, easy
removable tray for storage
|
|
Sample volume |
2 – 30 μL |
|
Sample tubes/cups |
Primary tubes of 5, 7 and 10 mL and sample cups |
|
Sample types |
Serum, plasma, urine, CSF, whole blood |
|
Reagent on-board capacity |
30 different methods
in bar coded mono - or twin - containers for adapter-free one-grip
loading
|
|
Reagent pipetting |
Reagent 1: 120 – 250 μL Reagent 2: 10 – 50 μL |
|
Reaction temperature |
37 ± 0.2 °C |
|
Reaction unit |
Temperature controlled rotating tray with 105 disposable plastic
cuvettes (37 ± 0.2 °C)
|
|
Photometry |
12 wavelengths: 340,
380, 405, 450, 480, 508, 546, 570, 600, 660, 700 and 800 nm (mono-
and bichromatic)
|
|
Photometric linearity and resolution
|
0 – 3.0 OD; resolution: 0.0001 O
|
|
Water consumption |
< 1 liter per hour |
|
System interface |
Analyzer - PC: USB connectivity bi-directional; CPU: Pentium IV or
higher
|
|
Power source/power consumption |
AC 110/220 V, 60/50 Hz; 300 VA excluding PC/printer/monitor
|
|
Dimensions |
60 cm (W) x 67 cm (D) x 60 cm (H) |
|
Weight |
Approximately 60 kg |
|
Respons® 910 order number DI