System type |
Floor stand random access clinical chemistry analyzer |
|
STAT-analytics |
STAT port enables easy STAT sample determination |
|
1200 tests/hour |
1200 tests/hour |
|
Photometry throughput |
800 tests/hour |
|
ISE throughput (Na, K, Cl) |
600 tests/hour |
|
Sample throughput rate |
Maximum 800 samples/hour |
|
Measuring principle |
Colorimetry (Rate/End Point); Turbidimetric
Immunoassay; ISE: simultaneous 3 item
measurement of Na, K, Cl |
|
Barcode identification |
Automatic barcode scan for reagents and samples |
|
Calibration |
Linear, non-linear, multi point |
|
Sample tray STT |
84 positions for routine and priority samples with
built-in barcode reader |
|
Refrigerated sample tray CTT |
61 positions for controls and calibrators |
|
Sample dispensing volume |
1
– 25 μL (0.1 μL step) |
|
Sample tubes/cups |
Most commonly used blood collection tubes and sample
cups |
|
Sample types |
Serum, plasma, urine, whole blood |
|
Sample dilution |
Dilution ratio: 2 to 300 times |
|
Sensors |
Liquid level-sensor and clot sensor |
|
Reagent dispensing method |
2 probe system capable to dispense 3 reagent
components |
|
Reagent trays |
2 trays for refrigerated 90 reagent containers with
built-in barcode reader |
|
Reagent tray standard setting |
Total 45 containers per tray: 12 containers with 70
mL and 33 containers with 40 mL |
|
Reagent container capacity |
20, 40, 70 mL with special adapters for 20 and 70 mL
containers |
|
Reagent pipetting |
5
to 300 μL |
|
Reaction disk |
Rotating reaction carousel with turntable and 231
reusable plastic cuvettes |
|
Reaction times and temperature |
3, 4, 5, 10, 15 minutes at 37 ± 0.1 °C; Extended
reaction time: 21 or 31 minutes |
|
Reflex testing capability |
Yes |
|
Photometry |
14
wavelengths: 340, 410, 451, 478, 505, 545, 571, 596, 658, 694, 751,
805, 845, 884 nm |
|
Automatic maintenance |
Automatic startup and shutdown by weekly timer |
|
Required pure water |
20 liter/hour; water purity: 1
μs/cm
or less |
|
Ambient temperature/humidity |
Temperatur: 18 °C to 30 °C/humidity: 40 % to 70 % |
|
Power supply |
AC
220 V ± 10 %, 50/60 Hz; 2.6 kVA |
|
Dimensions |
122 cm (W) x 85 cm (D) x 110.8 cm (H) |
|
Weight |
Approximately 450 kg |
|
BioMajesty JCA-BM6010/C order